This is also referred to the equipment being stuck in “Cooling Mode.” If your equipment is stuck in cooling mode it won’t go into ready mode.Į6 (Over-Current, Diode Issue): If this fault code presents itself on screen, this may require a reset by the operator which is done by turning the system off and waiting half a minute before restarting. Stuck in “Cooling Mode”: This issue may not display a fault code at all but this means that based on the results from the console’s energy detector, which measures the diode laser output, that it is too low. ![]() With a failed memory, the device will be unable to properly function and must be replaced before being able to use your LightSheer again. Unfortunately this fault code means that the embedded computer board is in need of replacement. ![]() Error/Fault Code “E3”: When operating one of the Lumenis-Lightsheer models, if this error message appears on screen, that means the device has a malfunctioning central processing unit (CPU).